APT3100 Smart Pressure Transmitter Operation Manual M3100-EO1G
30
Duon System Co.,Ltd.
(3) Intrinsical Safety Type
[Figure 4-3 Connection between Transmitter and Power Supply]
B. Wiring Installation
(1) General-use Type and intrinsically Safe Type
Make cable wiring using metallic conduit or
Waterproof cable glands.
o. Apply a non-hardening
sealant to the terminal
box connection port
and the threads on the
flexible metal conduit
for the waterproofing.
(2) KOSHA Flameproof Type
Wire cables through a flameproof packing adapter, or using a flameproof metal conduit.
(a) Wiring cable through flameproof packing adapter for KOSHA flameproof type (see Figure
4-4b)
Use only flameproof packing adapter by KOSHA.
Apply a non-hardening sealant to the terminal box connection port and to the threads
on the flameproof packing adapter for waterproofing
Mounting flameproof packing adapter to the terminal box.
o. Screw the flameproof packing adapter into the terminal box until the O-ring touches
the terminal box wiring port (at least 5 full turns), and tighten the lock net.
[Figure 4-4a
Typical Wiring using
Flexible Metal Conduit]