APT3100 Smart Pressure Transmitter Operation Manual M3100-EO1G
45
Duon System Co.,Ltd.
6.4.2.2 Fail Mode and Jumper Switch of EEPROM-write
Fail-mode and jumper switch of EEPROM-write is located front of electronics module(Refer to Figure
2-2, 2-3)
6.4.3 Assembling the Electronics Housing
Re-assembling procedure is same as follows.
1. Make sure that Fail-mode and Jumper Switch are set exactly.
2. Insert electronics module in housing
3. Firstly, connect the connector of sensor board with power connector.
z Mis connection of the two connectors cause wrong output(4~20mA) and effect on power
of Transmitter.
z In case of power connector sandwitched between board and housing, it may cause wrong
output signal and effect on power of transmitter.
4. Fix electronics module with 3 screws.
5. Close the cover of housing..
Figure 6.3 Structure of Electronics Module inner Transmitter
CPU &
Power Module
Analog
Connector
Power(24Vdc)
Connector